Study on solid phase extraction and graphite furnace atomic absorption spectrometry for the determination of nickel, silver, cobalt, copper, cadmium and lead with MCI GEL CHP 20Y as sorbent | |
Yang, Guangyu1,2,3; Fen, Weibo1,2; Lei, Chun1,2; Xiao, Weilie1,2; Sun, Handong1 | |
2009-02-15 | |
发表期刊 | JOURNAL OF HAZARDOUS MATERIALS |
ISSN | 0304-3894 |
卷号 | 162期号:1页码:44-49 |
摘要 | A solid phase extraction and graphite furnace atomic absorption spectrometry (GFAAS) for the determination of nickel, silver, cobalt, copper, cadmium and lead with MCI GEL CHP 20Y as sorbent was studied. Trace amounts of chromium, nickel, silver, cobalt, copper, cadmium and lead were reacted with 2-(2-quinolinil-azo)-4-methyl-1,3-dihydroxidobenzene (QAMDHB) followed by adsorption onto MCI GEL CHP 20Y solid phase extraction column, and 1.0 mol L(-1) HNO(3) Was used as eluent. The metal ions in 300 mL solution can be concentrated to 1.0 mL, representing an enrichment factor of 300 was achieved. The recoveries of analytes at pH 8.0 with 1.0 g of resin were greater than 95% without interference from alkaline, earth alkaline and some metal ions. When detected with graphite furnace atomic absorption spectrometry, the detection limits in the original samples were 1.4 ng L(-1) for Cr(III), 1.0 ng L(-1) for Ni(II), 0.85 ng L(-1) for Ag(I), 1.2 ng L(-1) for Co(II), 1.0 ng L(-1) for Cu(II), 1.2 ng L(-1) for Cd(II) and 1.3 ng L(-1) for Pb(II). The validation of the procedure was performed by the analysis of the certified standard reference materials, and the presented procedure was applied to the determination of analytes in biological, water and soil samples with good results (recoveries range from 89 to 104%, and R.S.D.% lower than 3.2%. The results agreed with the standard Value or reference method). (C) 2008 Published by Elsevier B.V. |
关键词 | Solid Phase Extraction Heavy Metal Ions Atomic Absorption Spectrometry 2-(2-quinolinil-azo)-4-methyl-1 Mci Gel Chp 20y Resin 3-dihydroxidobenzene |
学科领域 | Engineering, Environmental ; Engineering, Civil ; Environmental Sciences |
DOI | 10.1016/j.jhazmat.2008.05.007 |
收录类别 | SCI |
语种 | 英语 |
WOS记录号 | WOS:000263149400006 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.kib.ac.cn/handle/151853/2885 |
专题 | 植物化学与西部植物资源持续利用国家重点实验室 |
作者单位 | 1.Chinese Acad Sci, Kunming Inst Bot, State Key Lab Phytochem & Plant Resources W China, Kunming 650204, Peoples R China 2.Chinese Acad Sci, Grad Sch, Beijing 100039, Peoples R China 3.Yunnan Acad Tobacco Sci, Kunming 650106, Peoples R China |
推荐引用方式 GB/T 7714 | Yang, Guangyu,Fen, Weibo,Lei, Chun,et al. Study on solid phase extraction and graphite furnace atomic absorption spectrometry for the determination of nickel, silver, cobalt, copper, cadmium and lead with MCI GEL CHP 20Y as sorbent[J]. JOURNAL OF HAZARDOUS MATERIALS,2009,162(1):44-49. |
APA | Yang, Guangyu,Fen, Weibo,Lei, Chun,Xiao, Weilie,&Sun, Handong.(2009).Study on solid phase extraction and graphite furnace atomic absorption spectrometry for the determination of nickel, silver, cobalt, copper, cadmium and lead with MCI GEL CHP 20Y as sorbent.JOURNAL OF HAZARDOUS MATERIALS,162(1),44-49. |
MLA | Yang, Guangyu,et al."Study on solid phase extraction and graphite furnace atomic absorption spectrometry for the determination of nickel, silver, cobalt, copper, cadmium and lead with MCI GEL CHP 20Y as sorbent".JOURNAL OF HAZARDOUS MATERIALS 162.1(2009):44-49. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
201110260052.pdf(176KB) | 开放获取 | -- | 浏览 下载 |